The calibration plots showed a good linear relationship from 0.00800 mg/ml to 0.03200 mg/ml.
Validation of spectrophotometric dissolution method for modified release Trimetazidine pharmaceutical dosage form
Farzana Hasin, Md. Hasan Mukhlasy, Md. Iftekhar Hussain, Md. Mofazzal Hossain, Md. Mehedi Hasan
- Editor approved · 16 Nov 2026
- 2017
- DOI 10.56511/jipbs.2017.v4.i02.pp68-73.227
The question
Does a dissolution method for trimetazidine hydrochloride stand up to validation under ICH guidelines?
- The method was tested for linearity, accuracy and precision.
- The report states the dissolution conditions and that detection was at 231 ± 1 nm.
What was done
- Dissolution was run with USP II (Paddle), a phosphate buffer of pH 6.8 as the medium, and 75 rpm.
- Detection was at 231 ± 1 nm.
- Linearity was read from a linear regression analysis of the calibration plots.
- Linearity, accuracy and precision results were used to set the range of the method.
Key findings
A correlation coefficient of 1.00 was found.
About 95–105% of the trimetazidine hydrochloride was recovered.
The range of the method was established as 80%-120%, on the strength of the test results for linearity, accuracy and precision.
The method was validated in line with ICH guidelines.
Why it matters
- The report states that the method was validated as the ICH guidelines require.
- Results for linearity, accuracy and precision set the range of the method at 80%-120%.
How to cite
Farzana Hasin, Md. Hasan Mukhlasy, Md. Iftekhar Hussain, Md. Mofazzal Hossain, Md. Mehedi Hasan. Validation of spectrophotometric dissolution method for modified release Trimetazidine pharmaceutical dosage form. Journal of Innovations in Pharmaceutical and Biological Sciences. 2017;4(2):68-73. https://doi.org/10.56511/jipbs.2017.v4.i02.pp68-73.227
- The article at doi.org (the publisher's page and PDF)
- The article on the journal's site
- The journal at the ISSN Portal (ISSN 2349-2759)